Device Survival Probability

4092_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr                  
% Survival 98.8 98.7 98.5 98.1 97.8 97.4 97.4 97.4 97.4 97.4 97.4 97.4                  
#   921 822 734 629 515 402 333 279 239 157 93 57                  

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1202
Cumulative Months of Follow-Up 70183
Number of Leads Active in Study 12
Information 21

Qualifying Complications

Conductor Fracture 3
Extra Cardiac Stimulation 1
Failure to Capture 12
Impedance Out of Range 1
Lead Dislodgement 4
Other 0
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 4
Conductor Fracture 4
Extra Cardiac Stimulation 1
Failure to Capture 35
Impedance Out of Range 2
Insulation Breach 1
Lead Dislodgement 35
Oversensing 1
Unspecified Clinical Failure 1
Information

Distribution Data

US Market Release1998-09-17
CE Approval Date1998-04-15
Registered USA Implants186242
Estimated Active USA Implants36298
US Market Release1998-09-17
CE Approval Date1998-04-15
Estimated WW Distribution567957
Information

USA Returned Product Analysis

Conductor Fracture 21
Insulation Breach 99
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 21
Insulation Breach 107
Crimp/Weld/Bond 0
Other 2
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LEP
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of April 30, 2024

www.medtronic.com/productperformance 1001052