4074 CapSure Sense
Pacing
4074 CapSure Sense
VENTRICULAR PLACEMENT
Information

Device Survival Probability

4074_VEN_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr        
% Survival 99.4 99.3 99.2 99.2 99.2 99.2 99.2 99.2 99.2 98.8 98.8 97.5 97.5 96.7 96.7 96.7 96.7        
#   1029 880 739 633 488 395 338 294 258 212 175 153 125 117 111 88 58        

Product Surveillance Registry Results

Number of Leads Enrolled in Study 1193
Cumulative Months of Follow-Up 80631
Number of Leads Active in Study 158
Information 12

Qualifying Complications

Conductor Fracture 1
Failure to Capture 4
Impedance Out of Range 2
Insulation (not further defined) 2
Lead Dislodgement 2
Other 1
ATRIAL PLACEMENT
Information

Device Survival Probability

4074_ATR_SURV
Loading...











    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr 15 yr 16 yr 17 yr        
% Survival 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1 99.1        
#   214 205 198 183 167 158 148 136 126 117 110 107 99 95 88 77 56        

Product Surveillance Registry Results

Number of Leads Enrolled in Study 227
Cumulative Months of Follow-Up 29505
Number of Leads Active in Study 60
Information 2

Qualifying Complications

Failure to Capture 0
Failure to Sense 1
Impedance Out of Range 0
Lead Dislodgement 1
Other 0
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

US Acute Lead Observations

Cardiac Perforation 34
Conductor Fracture 2
Extra Cardiac Stimulation 3
Failure to Capture 190
Failure to Sense 17
Impedance Out of Range 9
Lead Dislodgement 208
Oversensing 9
Information

Distribution Data

US Market Release2002-06-23
CE Approval Date2002-02-01
Registered USA Implants154996
Estimated Active USA Implants74234
US Market Release2002-06-23
CE Approval Date2002-02-01
Estimated WW Distribution1086031
Information

USA Returned Product Analysis

Conductor Fracture 15
Insulation Breach 59
Crimp/Weld/Bond 0
Other 0
Information

WW Returned Product Analysis

Conductor Fracture 21
Insulation Breach 64
Crimp/Weld/Bond 0
Other 5
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix BBD
Lead Body Diameter 5.3 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of April 30, 2024