DVPA2D1 Cobalt XT
Implantable Cardioverter Defibrillator (ICD)
DVPA2D1 Cobalt XT
Information

Device Survival Probability

COBALT_CROME_VR_SURV
Loading...










    1 yr at 23.0 mo                
% Excluding Normal Battery Depletion 100.0 100.0                
% Including Normal Battery Depletion 100.0 100.0                
#   4032 199                
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 40 J
More
Serial Number Prefix RSC
Mass 79 g
Volume 33.2 cc
Waveform Biphasic
X-Ray ID PLS
Hide this content
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 5.2 5.3 5.4 5.5
Quarterly 8.7 9.1 9.5 9.6
Semi-annual 10.5 11.2 11.7 11.9
Footnote

Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm;Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 2V; Ventricle Impedance 500ohms; Ventricle Sensing Rate 70bpm; EGM ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation.Hide this content

Information

Distribution Data

US Market Release2020-04-23
CE Approval Date2019-12-18
Registered USA Implants770
Estimated Active USA Implants760
Normal Battery Depletions
US Market Release2020-04-23
CE Approval Date2019-12-18
Estimated WW Distribution1456
Normal Battery Depletions
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised
Therapy Function Compromised
Information

EOS Indicator

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.8 V on 3 consecutive daily automatic measurements.

Data as of July 27, 2022

www.medtronic.com/productperformance 876050