7232B Maximo VR
Implantable Cardioverter Defibrillator (ICD)
7232B Maximo VR
Information

Device Survival Probability

7232_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr at 161.0 mo              
% Excluding Normal Battery Depletion 100.0 99.9 99.9 99.9 99.8 99.8 99.8 99.8 99.8 99.8 99.8 99.8 99.8 99.8              
% Including Normal Battery Depletion 99.3 99.1 98.7 98.2 96.7 90.7 82.2 70.1 44.6 18.5 13.7 12.9 12.1 11.2              
#   37962 33964 30266 26674 23484 20403 17220 13744 8134 2797 1686 1216 599 131              
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 35J
More
Connector Style B Style
Serial Number Prefix PVF
Mass 76g
Volume 39cc
Waveform Biphasic
X-Ray ID PVF
Hide this content
B Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.4 4.7 4.8 4.9
Quarterly 7.0 7.5 8.0 8.3
Semi-annual 8.2 9.0 9.7 10.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2003-10-06
CE Approval Date2004-10-22
Registered USA Implants170
Estimated Active USA Implants22
Normal Battery Depletions34
US Market Release2003-10-06
CE Approval Date2004-10-22
Estimated WW Distribution217
Normal Battery Depletions34
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised
Therapy Function Compromised
Information

EOL Indication

EOL device status is defined as three months following an ERI indication assuming the following post-ERI conditions: 100% VVI pacing at 60 ppm, 3 V, 0.4 ms; 500 Ohm pacing load; and six 30 J charges. EOL may be indicated before the end of three months if the device exceeds these conditions.

Information

Elective Replacement (ERI)

Battery voltage less than or equal to 2.62v or charge time greater than 16 seconds.

Information

Charge Time

7232_CHRG
Loading...

Data as of November 8, 2019

www.medtronic.com/productperformance 726050