W4TR06
Solara Quad CRTP MRI SureScan

Device Survival Probability
W4TRXX_SURV
Loading...
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | |||||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
100.0 | 100.0 | 100.0 | 100.0 | 100.0 | |||||
% | ![]() |
100.0 | 99.9 | 99.8 | 99.2 | 98.9 | |||||
# | 36662 | 23629 | 13813 | 5696 | 233 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Product Characteristics
Generator Type | CRT-Pacing |
---|---|
NBD Code | OAE-DDDR |
Max. Delivered Energy | N/A |
Serial Number Prefix | RNZ |
---|---|
Mass | 30 g |
Volume | 20.5 cc |
Waveform | N/A |
X-Ray ID | RNP |

Estimated Longevity
Amplitude Setting | 500 Lead Ω | 1000 Lead Ω |
---|---|---|
Low 2.5 V (A, RV) | 8.7 | 10.9 |
Nominal 3.5 V (A, RV) | 6.9 | 8.5 |
High 5.0 V (A, RV) | 3.9 | 6.0 |
Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Left Ventricle Pulse Width 0.4 ms; Left Ventricle Pulse Amplitude same as A, RV; Left Ventricle Percent Paced 100 %; Left Ventricle Lead Impedance same as A, RV; EGM Prestorage ON ; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 % Hide this content

Distribution Data
US Market Release | |
---|---|
CE Approval Date | 2017-02-10 |
Registered USA Implants | |
Estimated Active USA Implants | |
Normal Battery Depletions |
US Market Release | |
---|---|
CE Approval Date | 2017-02-10 |
Estimated WW Distribution | 11803 |
Normal Battery Depletions |
Data as of July 27, 2022