W1TR04 Percepta CRTP MRI
Cardiac Resynchronization Therapy - Pacing (CRT-Pacing)
W1TR04 Percepta CRTP MRI
Information

Device Survival Probability

W1TRXX_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr at 51.0 mo          
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0          
% Including Normal Battery Depletion 99.9 99.9 99.6 97.6 96.7          
#   8854 5328 2364 331 126          
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type CRT-Pacing
NBD Code OAE-DDDR
Max. Delivered Energy N/A
More
Connector Style CRT-P Style
Serial Number Prefix RNU
Mass 30 g
Volume 20 cc
Waveform N/A
X-Ray ID RNP
Hide this content
CRT-P Style
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 8.7 10.9
Nominal 3.5 V (A, RV) 6.9 8.5
High 5.0 V (A, RV) 3.9 6.0
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Left Ventricle Pulse Width 0.4 ms; Left Ventricle Pulse Amplitude same as A, RV; Left Ventricle Percent Paced 100 %; Left Ventricle Lead Impedance same as A, RV; EGM Prestorage ON ; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 %Hide this content

Information

Distribution Data

US Market Release
CE Approval Date2017-02-10
Registered USA Implants
Estimated Active USA Implants
Normal Battery Depletions
US Market Release
CE Approval Date2017-02-10
Estimated WW Distribution7496
Normal Battery Depletions
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 2
Other 2
Therapy Function Compromised 1
Battery 1
Information

EOL Indicator

6 months after RRT

Information

Recommended Replacement (RRT)

180 days after 3 consecutive daily automatic measurements of battery voltage less than or equal to 2.63 V or immediately after 3 consecutive daily automatic measurements of less than or equal to 2.60 V, whichever comes first

Data as of December 16, 2021

www.medtronic.com/productperformance 851050