W1SR01
Azure XT SR

Device Survival Probability
AZURE_SR_SURV
Loading...
1 yr | 2 yr | at 33.0 mo | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
100.0 | 100.0 | 100.0 | |||||||
% | ![]() |
100.0 | 99.9 | 99.9 | |||||||
# | 14965 | 5088 | 203 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Product Characteristics
Generator Type | IPG |
---|---|
NBD Code | OOE-VVIR |
Max. Delivered Energy | N/A |
Connector Style | IS-1 UNI/BI |
---|---|
Serial Number Prefix | RNA |
Mass | 22.5 g |
Volume | 12.25 cc |
Waveform | N/A |
X-Ray ID | RNA |
IS-1 UNI/BI

Estimated Longevity
Amplitude Setting | 500 Lead Ω | 1000 Lead Ω |
---|---|---|
Low 2.5 V (A, RV) | 13.1 | 15.6 |
Nominal 3.5 V (A, RV) | 11.2 | 13.7 |
High 5.0 V (A, RV) | 8.2 | 11.0 |
Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; EGM Prestorage ON Time 1 monthHide this content
Data as of October 29, 2020