W1SR01 Azure XT SR
Implantable Pulse Generator (IPG)
W1SR01 Azure XT SR
Information

Device Survival Probability

AZURE_SR_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr at 69.0 mo        
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0        
% Including Normal Battery Depletion 100.0 100.0 99.9 99.9 99.8 99.5        
#   43364 30988 20185 11384 3919 246        
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code OOE-VVIR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix RNA
Mass 22.5 g
Volume 12.25 cc
Waveform N/A
X-Ray ID RNA
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 13.1 15.6
Nominal 3.5 V (A, RV) 11.2 13.7
High 5.0 V (A, RV) 8.2 11.0
Footnote

Longevity estimates based on the following device usage.   Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; EGM Prestorage ON Time 1 monthHide this content

Information

Distribution Data

US Market Release2017-08-16
CE Approval Date2017-03-02
Registered USA Implants50025
Estimated Active USA Implants41455
Normal Battery Depletions13
US Market Release2017-08-16
CE Approval Date2017-03-02
Estimated WW Distribution54197
Normal Battery Depletions13
Information

Malfunctions (USA)

Therapy Function Not Compromised 8
Battery 1
Electrical Component 4
Software/Firmware 1
Other 2
Therapy Function Compromised 1
Electrical Component 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 8
Battery 1
Electrical Component 4
Software/Firmware 1
Other 2
Therapy Function Compromised 1
Electrical Component 1
Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately six months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.63V on 3 consecutive daily automatic measurements.

Data as of December 1, 2023

www.medtronic.com/productperformance 951051