SSR203 Sigma 200 SR
Implantable Pulse Generator (IPG)
SSR203 Sigma 200 SR
Information

Device Survival Probability


SSR203_SS203_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr at 150 mo                
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 99.9 99.9 99.8 99.7 99.7 99.7 99.7 99.7 99.7                
% Including Normal Battery Depletion 99.9 99.9 99.8 99.7 99.2 98.6 97.2 95.2 92.3 86.6 76.2 65.6 59.6                
#   10350 8736 7434 6384 5485 4753 4061 3508 2829 2036 1245 457 110                
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PJG
Mass 21.5 g
Volume 10.5 cc
Waveform N/A
X-Ray ID PJG
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 10.1 11.1
Nominal 3.5 V (A, RV) 8.2 9.8
High 5.0 V (A, RV) 6.4 8.4
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppmHide this content

Information

EOL Indication

3 months after ERI.

Information

Distribution Data

US Market Release1999-09-02
CE Approval Date
Registered USA Implants12113
Estimated Active USA Implants1822
Normal Battery Depletions397
US Market Release1999-09-02
CE Approval Date
Estimated WW Distribution39677
Normal Battery Depletions740
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 13
Electrical Interconnect 13
Information

Malfunctions (WW)

Therapy Function Not Compromised 11
Electrical Component 1
Electrical Interconnect 7
Other 3
Therapy Function Compromised 71
Electrical Component 2
Electrical Interconnect 67
Other 2
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050