SS303 Sigma 300 S
Implantable Pulse Generator (IPG)
SS303 Sigma 300 S
Information

Device Survival Probability


SSR303_SSR306_SS303_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr at 146 mo                
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 99.9 99.8 99.7 99.7 99.7 99.6 99.6 99.6                
% Including Normal Battery Depletion 99.9 99.9 99.7 99.5 99.2 98.6 97.5 95.6 92.1 85.8 76.0 62.7 58.2                
#   47111 39971 34129 29348 25262 21010 16840 13027 9324 5638 2810 576 252                
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code SSI
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PJG
Mass 21.5 g
Volume 10.5 cc
Waveform N/A
X-Ray ID PJG
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 10.1 11.1
Nominal 3.5 V (A, RV) 8.2 9.8
High 5.0 V (A, RV) 6.4 8.4
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppmHide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-09-15
CE Approval Date1998-12-17
Registered USA Implants247
Estimated Active USA Implants61
Normal Battery Depletions
US Market Release1999-09-15
CE Approval Date1998-12-17
Estimated WW Distribution46138
Normal Battery Depletions87
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 7
Electrical Component 2
Electrical Interconnect 4
Other 1
Therapy Function Compromised 22
Electrical Component 1
Electrical Interconnect 20
Possible Early Battery Depletion 1
Information

Elective Replacement (ERI)

Replacement message on programmer. Nonmagnet mode VVI 65 ppm Magnet mode VOO 65 ppm

Data as of March 12, 2013