SS106 Sigma 100 S
Implantable Pulse Generator (IPG)
SS106 Sigma 100 S
Information

Device Survival Probability

SS103_SS106_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 104.0 mo                        
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0                        
% Including Normal Battery Depletion 100.0 100.0 99.8 99.8 99.0 99.0 97.9 96.4 91.2                        
#   600 473 371 294 225 189 154 122 100                        
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code SSI
Max. Delivered Energy N/A
More
Connector Style 5/6mm UNI
Serial Number Prefix PJH
Mass 21.6 g
Volume 10.8 cc
Waveform N/A
X-Ray ID PJH
Hide this content
5/6mm UNI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 10.1 11.1
Nominal 3.5 V (A, RV) 8.2 9.8
High 5.0 V (A, RV) 6.4 8.4
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppmHide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-08-30
CE Approval Date1998-12-17
Registered USA Implants68
Estimated Active USA Implants2
Normal Battery Depletions8
US Market Release1999-08-30
CE Approval Date1998-12-17
Estimated WW Distribution4233
Normal Battery Depletions93
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 1
Electrical Interconnect 1
Therapy Function Compromised 14
Electrical Interconnect 14
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of July 31, 2017

www.medtronic.com/productperformance 501050