SS103 Sigma 100 S
Implantable Pulse Generator (IPG)
SS103 Sigma 100 S
Information

Device Survival Probability


SS103_SS106_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 113 mo
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0
% Including Normal Battery Depletion 100.0 100.0 99.6 99.6 98.8 98.8 97.0 93.7 85.3 82.2
#   623 495 393 318 249 220 187 151 116 101
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code SSI
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PJG
Mass 21.5 g
Volume 10.5 cc
Waveform N/A
X-Ray ID PJG
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 10.1 11.1
Nominal 3.5 V (A, RV) 8.2 9.8
High 5.0 V (A, RV) 6.4 8.4
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppmHide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-08-30
CE Approval Date1998-12-17
Registered USA Implants761
Estimated Active USA Implants103
Normal Battery Depletions19
US Market Release1999-08-30
CE Approval Date1998-12-17
Estimated WW Distribution91132
Normal Battery Depletions91
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 5
Electrical Interconnect 5
Therapy Function Compromised 89
Electrical Component 4
Electrical Interconnect 83
Possible Early Battery Depletion 1
Other 1
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 pmm and VVI (VOO/65 with magnet).

Data as of March 12, 2013