SEDR01 Sensia DR
Implantable Pulse Generator (IPG)
SEDR01 Sensia DR
Information

Device Survival Probability

SEDR01_SED01_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr 14 yr at 179.0 mo            
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0            
% Including Normal Battery Depletion 99.9 99.9 99.8 99.6 99.2 98.6 97.3 93.2 84.0 71.3 58.1 47.2 41.3 39.2 38.5            
#   120560 109033 98382 88748 80022 72221 64926 54917 41774 28948 18360 10649 5778 2580 168            
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix NWL, PWL
Mass 27.1 g
Volume 12.1 cc
Waveform N/A
X-Ray ID PWL, PWB
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.4 | 9.1* 8.2 | 10.3*
Nominal 3.5 V (A, RV) 6.0 | 7.2* 7.4 | 9.0*
High 5.0 V (A, RV) 4.5 | 5.1* 6.0 | 7.0*
Footnote

*The second longevity numbers provided apply to the N-series Sensia SEDR01 product. The N-series products have increased longevity because of improved electronic circuitry. Click on "Other Articles" for more information on the N-series longevity. Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; EGM Prestorage ON Time 1 ; Atrial Fib Percent of Time 0 %Hide this content

Information

Distribution Data

US Market Release2006-07-17
CE Approval Date2005-09-20
Registered USA Implants149401
Estimated Active USA Implants31023
Normal Battery Depletions15622
US Market Release2006-07-17
CE Approval Date2005-09-20
Estimated WW Distribution663745
Normal Battery Depletions16085
Information

Malfunctions (USA)

Therapy Function Not Compromised 17
Electrical Component 15
Electrical Interconnect 1
Other 1
Therapy Function Compromised 16
Electrical Component 6
Electrical Interconnect 3
Possible Early Battery Depletion 1
Other 6
Information

Malfunctions (WW)

Therapy Function Not Compromised 51
Electrical Component 44
Electrical Interconnect 1
Possible Early Battery Depletion 2
Other 4
Therapy Function Compromised 26
Electrical Component 15
Electrical Interconnect 4
Possible Early Battery Depletion 1
Other 6
Information

EOS Indication

The prolonged service period is 3 months after RRT/ERI and assumes 100% pacing at VVI/65 and the listed parameter values. Reprogramming of parameters may shorten the duration of the prolonged service period.

Information

Recommended Replacement (RRT)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of December 1, 2023

www.medtronic.com/productperformance 951051