Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
Excluding Normal Battery Depletion – This is the malfunction free survival curve.
*The second longevity numbers provided apply to the N-series Sensia SED01 product. The N-series products have increased longevity because of improved electronic circuitry. Click on "Other Articles" for more information on the N-series longevity.
Longevity estimates based on the following device usage.
Pace/Sense Mode DDD ;
Atrial Pulse Width 0.4 ms;
Right Ventricle Pulse Width 0.4 ms;
Lower Pace Rate 60 bpm;
Right Ventricle Percent Paced 100 %;
EGM Prestorage ON ;
Upper Sensing Rate 130 ppm;
Atrial Percent Paced 100 %;
EGM Prestorage ON Time 1 ;
Atrial Fib Percent of Time 0 %Hide this content
Distribution Data
US Market Release
2006-07-17
CE Approval Date
2005-09-20
Registered USA Implants
5
Estimated Active USA Implants
1
Normal Battery Depletions
1
US Market Release
2006-07-17
CE Approval Date
2005-09-20
Estimated WW Distribution
65565
Normal Battery Depletions
48
Malfunctions (USA)
Therapy Function Not Compromised
Therapy Function Compromised
Malfunctions (WW)
Therapy Function Not Compromised
4
Electrical Component
4
Therapy Function Compromised
3
Electrical Component
3
EOS Indication
Prolonged service period is 3 months after RRT/ERI assuming 100% VVI 65 ppm pacing at the listed parameter settings. Reprogramming of parameters can reduce the prolonged service period.
Recommended Replacement (RRT)
Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).