SDR306 Sigma 300 DR
Implantable Pulse Generator (IPG)
SDR306 Sigma 300 DR
Information

Device Survival Probability


SDR303_SDR306_SD303_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 143 mo                  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 99.9 99.9 99.9 99.7 99.6 99.5 99.4 99.4 99.4                  
% Including Normal Battery Depletion 99.9 99.9 99.8 99.6 99.3 98.8 97.7 95.1 91.0 83.6 69.4 39.5                  
#   96607 86521 77439 69038 61285 52984 43476 33528 23987 14385 6341 476                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD/RO
Max. Delivered Energy N/A
More
Connector Style 5/6mm UNI
Serial Number Prefix PJE
Mass 27.0 g
Volume 15.9 cc
Waveform N/A
X-Ray ID PJE
Hide this content
5/6mm UNI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 10.1 11.7
Nominal 3.5 V (A, RV) 7.5 9.6
High 5.0 V (A, RV) 5.5 7.8
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-08-30
CE Approval Date1998-12-17
Registered USA Implants1207
Estimated Active USA Implants203
Normal Battery Depletions116
US Market Release1999-08-30
CE Approval Date1998-12-17
Estimated WW Distribution2279
Normal Battery Depletions126
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 5
Electrical Interconnect 5
Information

Malfunctions (WW)

Therapy Function Not Compromised 2
Electrical Interconnect 2
Therapy Function Compromised 6
Electrical Interconnect 6
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050