SDR303 Sigma 300 DR
Implantable Pulse Generator (IPG)
SDR303 Sigma 300 DR
Information

Device Survival Probability


SDR303_SDR306_SD303_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 143 mo                  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 99.9 99.9 99.9 99.7 99.6 99.5 99.4 99.4 99.4                  
% Including Normal Battery Depletion 99.9 99.9 99.8 99.6 99.3 98.8 97.7 95.1 91.0 83.6 69.4 39.5                  
#   96607 86521 77439 69038 61285 52984 43476 33528 23987 14385 6341 476                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD/RO
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PJD
Mass 27.2 g
Volume 12.8 cc
Waveform N/A
X-Ray ID PJD
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 10.1 11.7
Nominal 3.5 V (A, RV) 7.5 9.6
High 5.0 V (A, RV) 5.5 7.8
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-08-26
CE Approval Date1998-12-17
Registered USA Implants105277
Estimated Active USA Implants29760
Normal Battery Depletions3754
US Market Release1999-08-26
CE Approval Date1998-12-17
Estimated WW Distribution190381
Normal Battery Depletions4113
Information

Malfunctions (USA)

Therapy Function Not Compromised 58
Electrical Component 6
Electrical Interconnect 41
Possible Early Battery Depletion 1
Other 10
Therapy Function Compromised 181
Electrical Component 7
Electrical Interconnect 173
Other 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 115
Electrical Component 11
Electrical Interconnect 88
Possible Early Battery Depletion 1
Software/Firmware 1
Other 14
Therapy Function Compromised 350
Electrical Component 11
Electrical Interconnect 338
Other 1
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013