SDR203 Sigma 200 DR
Implantable Pulse Generator (IPG)
SDR203 Sigma 200 DR
Information

Device Survival Probability


SDR203_SD203_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr at 145 mo                
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 99.9 99.9 99.8 99.6 99.5 99.5 99.5 99.5 99.5                
% Including Normal Battery Depletion 100.0 99.9 99.9 99.7 99.4 98.8 97.3 94.3 90.5 83.5 69.4 48.8 42.6                
#   14198 12732 11324 10143 9008 7971 6933 5861 4673 3232 1674 259 172                
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PJD
Mass 27.2 g
Volume 12.8 cc
Waveform N/A
X-Ray ID PJD
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 10.1 11.7
Nominal 3.5 V (A, RV) 7.5 9.6
High 5.0 V (A, RV) 5.5 7.8
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-08-31
CE Approval Date1998-12-17
Registered USA Implants15635
Estimated Active USA Implants3261
Normal Battery Depletions812
US Market Release1999-08-31
CE Approval Date1998-12-17
Estimated WW Distribution38933
Normal Battery Depletions1024
Information

Malfunctions (USA)

Therapy Function Not Compromised 12
Electrical Component 1
Electrical Interconnect 10
Other 1
Therapy Function Compromised 23
Electrical Component 1
Electrical Interconnect 21
Other 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 33
Electrical Component 5
Electrical Interconnect 20
Software/Firmware 3
Other 5
Therapy Function Compromised 84
Electrical Component 1
Electrical Interconnect 82
Other 1
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050