SD203 Sigma 200 D
Implantable Pulse Generator (IPG)
SD203 Sigma 200 D
Information

Device Survival Probability

SDR203_SD203_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr at 164.0 mo              
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 99.9 99.9 99.8 99.6 99.5 99.4 99.4 99.2 99.0 99.0              
% Including Normal Battery Depletion 99.7 99.5 99.4 99.1 98.6 97.6 95.4 92.0 86.8 78.2 64.5 47.2 31.1 20.6              
#   12594 11277 9975 8830 7720 6721 5724 4835 4035 3201 2218 1211 469 128              
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PJD
Mass 27.2 g
Volume 12.8 cc
Waveform N/A
X-Ray ID PJD
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 10.1 11.7
Nominal 3.5 V (A, RV) 7.5 9.6
High 5.0 V (A, RV) 5.5 7.8
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-08-31
CE Approval Date1998-12-17
Registered USA Implants228
Estimated Active USA Implants18
Normal Battery Depletions18
US Market Release1999-08-31
CE Approval Date1998-12-17
Estimated WW Distribution21852
Normal Battery Depletions269
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 1
Electrical Interconnect 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 7
Electrical Component 1
Electrical Interconnect 5
Other 1
Therapy Function Compromised 87
Electrical Component 2
Electrical Interconnect 85
Information

Elective Replacement (ERI)

Replacement message on programmer. Nonmagnet mode VVI 65 ppm Magnet mode VOO 65 ppm

Data as of November 2, 2015

www.medtronic.com/productperformance 376051