RVDR01 Revo MRI SureScan
Implantable Pulse Generator (IPG)
RVDR01 Revo MRI SureScan
Information

Device Survival Probability

RVDR01_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 83.0 mo      
% Excluding Normal Battery Depletion 100.0 100.0 100.0 99.9 99.9 99.9 99.8      
% Including Normal Battery Depletion 100.0 99.9 99.9 99.8 99.6 99.2 93.7      
#   60993 57558 54051 50468 42936 20136 504      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDRP
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PTN
Mass 21.5 g
Volume 12.7 cc
Waveform N/A
X-Ray ID PTA and MRI symbol
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 9.7 11.2
Nominal 3.5 V (A, RV) 7.3 9.4
High 5.0 V (A, RV) 4.9 7.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

Distribution Data

US Market Release2011-02-08
CE Approval Date
Registered USA Implants68905
Estimated Active USA Implants58026
Normal Battery Depletions225
US Market Release2011-02-08
CE Approval Date
Estimated WW Distribution69610
Normal Battery Depletions225
Information

Malfunctions (USA)

Therapy Function Not Compromised 61
Battery 1
Electrical Component 32
Possible Early Battery Depletion 24
Software/Firmware 3
Other 1
Therapy Function Compromised 3
Electrical Component 3
Information

Malfunctions (WW)

Therapy Function Not Compromised 63
Battery 1
Electrical Component 34
Possible Early Battery Depletion 24
Software/Firmware 3
Other 1
Therapy Function Compromised 3
Electrical Component 3
Information

EOS Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of January 31, 2018