RESR01 Relia SR
Implantable Pulse Generator (IPG)
RESR01 Relia SR
Information

Device Survival Probability

ADSR01_ADSR03_ADSR06_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 135.0 mo                  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0                  
% Including Normal Battery Depletion 99.9 99.7 99.5 99.0 98.3 97.2 94.7 88.7 71.7 50.5 27.5 14.6                  
#   72024 62938 55043 47920 41117 34801 28762 21871 12930 5487 876 116                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code AAIR/VVIR, AAI/VVI
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix NWU
Mass 21.5 g
Volume 9.7 cc
Waveform N/A
X-Ray ID PWB
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 8.9 9.7
Nominal 3.5 V (A, RV) 7.7 9.1
High 5.0 V (A, RV) 5.9 7.4
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; EGM Prestorage ON ; Upper Sensing Rate 130 ppm; EGM Prestorage ON Time 1 monthHide this content

Information

Distribution Data

US Market Release
CE Approval Date2008-05-07
Registered USA Implants6
Estimated Active USA Implants0
Normal Battery Depletions
US Market Release
CE Approval Date2008-05-07
Estimated WW Distribution199171
Normal Battery Depletions70
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 8
Electrical Component 4
Possible Early Battery Depletion 2
Other 2
Therapy Function Compromised 5
Electrical Component 5
Information

EOS Indication

The prolonged service period is 3 months after RRT/ERI and assumes 100% pacing at VVI/65 and the listed parameter values. Reprogramming of parameters may shorten the duration of the prolonged service period.

Information

Recommended Replacement (RRT)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of December 1, 2023

www.medtronic.com/productperformance 951051