P1501DR EnRhythm DR
Implantable Pulse Generator (IPG)
P1501DR EnRhythm DR
Information

Device Survival Probability


P1501DR_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr at 86 mo    
% Excluding Normal Battery Depletion 100.0 100.0 99.9 99.1 96.2 89.4 85.4 85.0    
% Including Normal Battery Depletion 99.9 99.9 99.6 97.9 91.8 76.8 61.1 54.6    
#   103946 94221 78341 61642 44520 25132 4460 1301    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDRP
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PNP
Mass 21.0 g
Volume 13.0 cc
Waveform N/A
X-Ray ID PNP
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 9.7 11.2
Nominal 3.5 V (A, RV) 7.3 9.4
High 5.0 V (A, RV) 4.9 7.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %; These estimates are adjusted to reflect changes addressed by February 2010 Advisory "Low Battery Voltage Displayed at Device Interrogation".Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2005-05-05
CE Approval Date2004-08-13
Registered USA Implants110100
Estimated Active USA Implants67332
Normal Battery Depletions1013
US Market Release2005-05-05
CE Approval Date2004-08-13
Estimated WW Distribution142191
Normal Battery Depletions1184
Information

Malfunctions (USA)

Therapy Function Not Compromised 5897
Battery 5763
Electrical Component 38
Electrical Interconnect 2
Possible Early Battery Depletion 37
Other 57
Therapy Function Compromised 53
Battery 5
Electrical Component 37
Electrical Interconnect 4
Possible Early Battery Depletion 2
Other 5
Information

Malfunctions (WW)

Therapy Function Not Compromised 6831
Battery 6626
Electrical Component 72
Electrical Interconnect 5
Possible Early Battery Depletion 48
Software/Firmware 1
Other 79
Therapy Function Compromised 73
Battery 7
Electrical Component 51
Electrical Interconnect 5
Possible Early Battery Depletion 5
Other 5
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050