KSR906 Kappa 900 SR
Implantable Pulse Generator (IPG)
KSR906 Kappa 900 SR
Information

Device Survival Probability

KSR901_KSR903_KSR906_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr at 129.0 mo                    
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 99.9 99.9 99.9 99.9 99.9 99.9                    
% Including Normal Battery Depletion 99.5 99.1 98.8 97.7 96.2 93.1 83.3 59.6 36.6 26.9 23.9                    
#   31945 27507 23902 20586 17669 14997 11832 7300 2856 890 203                    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Performance Notes For This Model

    • There are no performance notes for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style 5/6mm UNI
Serial Number Prefix PLH
Mass 22.5 g
Volume 11.0 cc
Waveform N/A
X-Ray ID PLH
Hide this content
5/6mm UNI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.3 7.9
Nominal 3.5 V (A, RV) 6.4 7.4
High 5.0 V (A, RV) 4.9 6.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppmHide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2002-01-09
CE Approval Date2001-09-28
Registered USA Implants1322
Estimated Active USA Implants125
Normal Battery Depletions179
US Market Release2002-01-09
CE Approval Date2001-09-28
Estimated WW Distribution3462
Normal Battery Depletions227
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 1
Electrical Interconnect 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 0
Therapy Function Compromised 2
Electrical Interconnect 2
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of February 9, 2015

www.medtronic.com/productperformance 351052