KSR706 Kappa 700 SR
Implantable Pulse Generator (IPG)
KSR706 Kappa 700 SR
Information

Device Survival Probability

KSR700_KSR701_KSR703_KSR706_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 106.0 mo                        
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 99.9 99.9 99.8                        
% Including Normal Battery Depletion 99.5 99.1 98.4 96.8 93.7 88.0 73.8 42.8 5.3                        
#   41395 35433 29986 25068 20413 16037 11092 4962 467                        
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style 5/6mm UNI
Serial Number Prefix PHU
Mass 22.5 g
Volume 11.0 cc
Waveform N/A
X-Ray ID PHU
Hide this content
5/6mm UNI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.4 7.9
Nominal 3.5 V (A, RV) 6.5 7.5
High 5.0 V (A, RV) 4.9 6.2
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppmHide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-02-09
CE Approval Date1998-03-20
Registered USA Implants2921
Estimated Active USA Implants158
Normal Battery Depletions302
US Market Release1999-02-09
CE Approval Date1998-03-20
Estimated WW Distribution7296
Normal Battery Depletions393
Information

Malfunctions (USA)

Therapy Function Not Compromised 1
Electrical Component 1
Therapy Function Compromised 1
Electrical Component 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 2
Electrical Component 2
Therapy Function Compromised 7
Electrical Component 1
Electrical Interconnect 6
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of April 14, 2016

www.medtronic.com/productperformance 426050