KSR703 Kappa 700 SR
Implantable Pulse Generator (IPG)
KSR703 Kappa 700 SR
Information

Device Survival Probability

KSR700_KSR701_KSR703_KSR706_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 117.0 mo
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 99.9 99.9 99.8 99.8
% Including Normal Battery Depletion 99.8 99.7 99.3 98.5 96.9 93.4 84.0 61.4 29.8 8.9
#   48284 41647 35741 30631 25938 21432 16060 8258 1902 115
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PHW
Mass 22.5 g
Volume 10.5 cc
Waveform N/A
X-Ray ID PHW
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.4 7.9
Nominal 3.5 V (A, RV) 6.5 7.5
High 5.0 V (A, RV) 4.9 6.2
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppmHide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-02-08
CE Approval Date1998-03-20
Registered USA Implants3607
Estimated Active USA Implants288
Normal Battery Depletions389
US Market Release1999-02-08
CE Approval Date1998-03-20
Estimated WW Distribution45824
Normal Battery Depletions1242
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 4
Electrical Component 1
Electrical Interconnect 3
Information

Malfunctions (WW)

Therapy Function Not Compromised 5
Electrical Component 4
Possible Early Battery Depletion 1
Therapy Function Compromised 33
Electrical Component 2
Electrical Interconnect 31
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of January 31, 2014

www.medtronic.com/productperformance 301050