KSR701 Kappa 700 SR
Implantable Pulse Generator (IPG)
Advisories For This Model
Performance Notes For This Model
Product Characteristics
| Generator Type |
IPG |
| NBD Code |
SSIR |
| Max. Delivered Energy |
N/A |
More
| Connector Style |
IS-1 UNI/BI |
| Serial Number Prefix |
PHT |
| Mass |
21.5 g |
| Volume |
9.7 cc |
| Waveform |
N/A |
| X-Ray ID |
PHT |
Hide this content
Estimated Longevity
| Amplitude Setting |
500 Lead Ω |
1000 Lead Ω |
| Low 2.5 V (A, RV) |
7.4 |
7.9 |
| Nominal 3.5 V (A, RV) |
6.5 |
7.5 |
| High 5.0 V (A, RV) |
4.9 |
6.2 |
EOL Indication
From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.
Elective Replacement (ERI)
Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).
Data as of March 12, 2013
www.medtronic.com/productperformance
251050