KSR701 Kappa 700 SR
Implantable Pulse Generator (IPG)
KSR701 Kappa 700 SR
Information

Device Survival Probability

KSR700_KSR701_KSR703_KSR706_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 106.0 mo                        
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 99.9 99.9 99.8                        
% Including Normal Battery Depletion 99.5 99.1 98.4 96.8 93.8 88.1 74.0 43.0 5.7                        
#   41367 35421 29991 25077 20428 16066 11132 4997 482                        
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PHT
Mass 21.5 g
Volume 9.7 cc
Waveform N/A
X-Ray ID PHT
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.4 7.9
Nominal 3.5 V (A, RV) 6.5 7.5
High 5.0 V (A, RV) 4.9 6.2
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppmHide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-01-29
CE Approval Date1998-03-20
Registered USA Implants48459
Estimated Active USA Implants2786
Normal Battery Depletions5125
US Market Release1999-01-29
CE Approval Date1998-03-20
Estimated WW Distribution90112
Normal Battery Depletions5741
Information

Malfunctions (USA)

Therapy Function Not Compromised 3
Electrical Component 1
Electrical Interconnect 1
Possible Early Battery Depletion 1
Therapy Function Compromised 19
Electrical Component 2
Electrical Interconnect 17
Information

Malfunctions (WW)

Therapy Function Not Compromised 17
Electrical Component 10
Electrical Interconnect 4
Possible Early Battery Depletion 2
Software/Firmware 1
Therapy Function Compromised 73
Electrical Component 4
Electrical Interconnect 69
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of November 2, 2015

www.medtronic.com/productperformance 376051