KSR403 Kappa 400 SR
Implantable Pulse Generator (IPG)
KSR403 Kappa 400 SR
Information

Device Survival Probability


KSR401_KSR403_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 113 mo
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0
% Including Normal Battery Depletion 99.9 99.7 99.5 99.0 98.4 96.7 90.4 66.2 29.6 14.1
#   13613 11960 10448 9155 7908 6667 5121 2579 535 151
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code SSIR
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PGD
Mass 28.0 g
Volume 11.7 cc
Waveform N/A
X-Ray ID PGD
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.9 8.4
Nominal 3.5 V (A, RV) 6.9 7.7
High 5.0 V (A, RV) 5.8 7.0
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode VVI ; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Minute Ventilation ONHide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1998-02-24
CE Approval Date1996-11-12
Registered USA Implants3611
Estimated Active USA Implants494
Normal Battery Depletions327
US Market Release1998-02-24
CE Approval Date1996-11-12
Estimated WW Distribution6038
Normal Battery Depletions521
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 1
Electrical Interconnect 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 0
Therapy Function Compromised 2
Electrical Interconnect 2
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013