KDR921 Kappa 900 DR
Implantable Pulse Generator (IPG)
KDR921 Kappa 900 DR
Information

Device Survival Probability

KDR921_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 83.0 mo      
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0      
% Including Normal Battery Depletion 99.9 99.5 98.9 96.8 89.6 59.4 2.7      
#   14246 12704 11245 9713 7708 3511 145      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PKR
Mass 23.6 g
Volume 11.1 cc
Waveform N/A
X-Ray ID PKR
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 5.5 6.1
Nominal 3.5 V (A, RV) 4.4 5.5
High 5.0 V (A, RV) 3.0 4.3
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2002-01-09
CE Approval Date2001-09-28
Registered USA Implants16329
Estimated Active USA Implants1374
Normal Battery Depletions2843
US Market Release2002-01-09
CE Approval Date2001-09-28
Estimated WW Distribution22187
Normal Battery Depletions2880
Information

Malfunctions (USA)

Therapy Function Not Compromised 1
Electrical Component 1
Therapy Function Compromised 3
Electrical Interconnect 3
Information

Malfunctions (WW)

Therapy Function Not Compromised 2
Electrical Component 2
Therapy Function Compromised 3
Electrical Interconnect 3
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of January 31, 2014

www.medtronic.com/productperformance 301050