KDR721 Kappa 700 DR
Implantable Pulse Generator (IPG)
KDR721 Kappa 700 DR
Information

Device Survival Probability


KDR721_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr at 86 mo    
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 99.9 99.9 99.9 99.9    
% Including Normal Battery Depletion 99.9 99.5 98.7 96.6 90.7 68.9 22.2 12.3    
#   8630 7619 6647 5642 4453 2258 317 125    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD/RO
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PGR
Mass 23.6 g
Volume 11.1 cc
Waveform N/A
X-Ray ID PGR
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 5.5 6.1
Nominal 3.5 V (A, RV) 4.4 5.5
High 5.0 V (A, RV) 3.0 4.2
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-02-11
CE Approval Date1998-03-20
Registered USA Implants9830
Estimated Active USA Implants753
Normal Battery Depletions1346
US Market Release1999-02-11
CE Approval Date1998-03-20
Estimated WW Distribution35155
Normal Battery Depletions1405
Information

Malfunctions (USA)

Therapy Function Not Compromised 1
Electrical Component 1
Therapy Function Compromised 4
Electrical Interconnect 4
Information

Malfunctions (WW)

Therapy Function Not Compromised 5
Electrical Component 3
Possible Early Battery Depletion 1
Other 1
Therapy Function Compromised 29
Electrical Component 3
Electrical Interconnect 26
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of March 12, 2013