KDR706 Kappa 700 DR
Implantable Pulse Generator (IPG)
KDR706 Kappa 700 DR
Information

Device Survival Probability


KD700_KDR700_KDR701_KDR703_KDR706_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr  
% Excluding Normal Battery Depletion 100.0 99.9 99.9 99.9 99.9 99.9 99.7 99.6 99.4  
% Including Normal Battery Depletion 99.9 99.8 99.6 99.1 98.0 95.4 85.9 59.3 10.9  
#   180496 165516 151007 136550 122426 106966 81763 40586 3056  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD/RO
Max. Delivered Energy N/A
More
Connector Style 5/6mm UNI
Serial Number Prefix PGW
Mass 28.5 g
Volume 14.2 cc
Waveform N/A
X-Ray ID PGW
Hide this content
5/6mm UNI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.7 8.6
Nominal 3.5 V (A, RV) 6.3 7.7
High 5.0 V (A, RV) 4.4 6.0
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-02-09
CE Approval Date1998-03-20
Registered USA Implants2640
Estimated Active USA Implants188
Normal Battery Depletions396
US Market Release1999-02-09
CE Approval Date1998-03-20
Estimated WW Distribution4783
Normal Battery Depletions467
Information

Malfunctions (USA)

Therapy Function Not Compromised 1
Electrical Interconnect 1
Therapy Function Compromised 9
Electrical Interconnect 9
Information

Malfunctions (WW)

Therapy Function Not Compromised 2
Electrical Component 1
Electrical Interconnect 1
Therapy Function Compromised 27
Electrical Interconnect 27
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050