KDR703 Kappa 700 DR
Implantable Pulse Generator (IPG)
KDR703 Kappa 700 DR
Information

Device Survival Probability


KD700_KDR700_KDR701_KDR703_KDR706_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 109 mo
% Excluding Normal Battery Depletion 100.0 100.0 99.9 99.9 99.9 99.9 99.7 99.6 99.4 99.4
% Including Normal Battery Depletion 99.9 99.8 99.5 99.1 98.0 95.3 85.8 59.3 13.8 6.3
#   180525 165538 151029 136576 122438 106991 83675 43045 4419 2245
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD/RO
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PGY
Mass 28.1 g
Volume 13.0 cc
Waveform N/A
X-Ray ID PGY
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.7 8.6
Nominal 3.5 V (A, RV) 6.3 7.7
High 5.0 V (A, RV) 4.4 6.0
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-02-05
CE Approval Date1998-03-20
Registered USA Implants9226
Estimated Active USA Implants793
Normal Battery Depletions1506
US Market Release1999-02-05
CE Approval Date1998-03-20
Estimated WW Distribution55807
Normal Battery Depletions1599
Information

Malfunctions (USA)

Therapy Function Not Compromised 4
Electrical Component 3
Possible Early Battery Depletion 1
Therapy Function Compromised 30
Electrical Component 1
Electrical Interconnect 29
Information

Malfunctions (WW)

Therapy Function Not Compromised 9
Electrical Component 6
Possible Early Battery Depletion 2
Software/Firmware 1
Therapy Function Compromised 55
Electrical Component 3
Electrical Interconnect 52
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of January 31, 2014

www.medtronic.com/productperformance 301050