KDR701 Kappa 700 DR
Implantable Pulse Generator (IPG)
KDR701 Kappa 700 DR
Information

Device Survival Probability

KD700_KDR700_KDR701_KDR703_KDR706_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 112.0 mo
% Excluding Normal Battery Depletion 100.0 99.9 99.9 99.9 99.9 99.8 99.7 99.6 99.4 99.4
% Including Normal Battery Depletion 99.4 99.2 98.8 97.9 96.2 92.3 80.1 51.1 14.9 2.6
#   167645 152726 138235 123727 109556 94147 72300 38576 8048 1729
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD/RO
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PGU
Mass 27.1 g
Volume 12.1 cc
Waveform NA
X-Ray ID PGU
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.7 8.6
Nominal 3.5 V (A, RV) 6.3 7.7
High 5.0 V (A, RV) 4.4 6.0
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-01-29
CE Approval Date1998-03-20
Registered USA Implants194278
Estimated Active USA Implants11072
Normal Battery Depletions37229
US Market Release1999-01-29
CE Approval Date1998-03-20
Estimated WW Distribution272287
Normal Battery Depletions37693
Information

Malfunctions (USA)

Therapy Function Not Compromised 46
Battery 1
Electrical Component 23
Electrical Interconnect 18
Possible Early Battery Depletion 3
Other 1
Therapy Function Compromised 653
Electrical Component 16
Electrical Interconnect 636
Possible Early Battery Depletion 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 69
Battery 1
Electrical Component 29
Electrical Interconnect 29
Possible Early Battery Depletion 5
Software/Firmware 3
Other 2
Therapy Function Compromised 986
Battery 1
Electrical Component 28
Electrical Interconnect 956
Possible Early Battery Depletion 1
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of February 3, 2017