KDR653 Kappa 600 DR
Implantable Pulse Generator (IPG)
KDR653 Kappa 600 DR
Information

Device Survival Probability


KDR651_KDR653_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 103 mo  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 99.9 99.7 99.5 99.4  
% Including Normal Battery Depletion 99.9 99.9 99.8 99.3 98.1 94.7 80.8 43.2 5.5  
#   12573 11383 10297 9173 8090 6959 4949 1834 163  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD/RO
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PLK
Mass 28.1 g
Volume 13.0 cc
Waveform N/A
X-Ray ID PLJ
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.7 8.6
Nominal 3.5 V (A, RV) 6.3 7.7
High 5.0 V (A, RV) 4.4 6.0
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2001-03-08
CE Approval Date
Registered USA Implants663
Estimated Active USA Implants29
Normal Battery Depletions113
US Market Release2001-03-08
CE Approval Date
Estimated WW Distribution687
Normal Battery Depletions113
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 10
Electrical Interconnect 10
Information

Malfunctions (WW)

Therapy Function Not Compromised 0
Therapy Function Compromised 10
Electrical Interconnect 10
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050