KDR651 Kappa 600 DR
Implantable Pulse Generator (IPG)
KDR651 Kappa 600 DR
Information

Device Survival Probability


KDR651_KDR653_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 103 mo  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 99.9 99.7 99.5 99.4  
% Including Normal Battery Depletion 99.9 99.9 99.8 99.3 98.1 94.7 80.8 43.2 5.5  
#   12573 11383 10297 9173 8090 6959 4949 1834 163  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD/RO
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PLJ
Mass 27.1 g
Volume 12.1 cc
Waveform N/A
X-Ray ID PLJ
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.7 8.6
Nominal 3.5 V (A, RV) 6.3 7.7
High 5.0 V (A, RV) 4.4 6.0
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2001-03-08
CE Approval Date
Registered USA Implants15689
Estimated Active USA Implants1196
Normal Battery Depletions2656
US Market Release2001-03-08
CE Approval Date
Estimated WW Distribution15982
Normal Battery Depletions2661
Information

Malfunctions (USA)

Therapy Function Not Compromised 3
Electrical Component 1
Possible Early Battery Depletion 2
Therapy Function Compromised 44
Electrical Component 1
Electrical Interconnect 42
Other 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 5
Electrical Component 2
Possible Early Battery Depletion 2
Other 1
Therapy Function Compromised 44
Electrical Component 1
Electrical Interconnect 42
Other 1
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050