KDR603 Kappa 600 DR
Implantable Pulse Generator (IPG)
KDR603 Kappa 600 DR
Information

Device Survival Probability


KDR600_KDR601_KDR603_KDR606_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 99.9 99.9 99.9 99.8 99.7 99.6  
% Including Normal Battery Depletion 99.9 99.8 99.8 99.4 98.7 96.8 87.7 58.4 2.7  
#   21342 19411 17444 15602 13810 11935 9120 4356 174  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD/RO
Max. Delivered Energy N/A
More
Connector Style 3.2mm Long Pin
Serial Number Prefix PHH
Mass 28.1 g
Volume 13.0 cc
Waveform N/A
X-Ray ID PHH
Hide this content
3.2mm Long Pin
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.7 8.6
Nominal 3.5 V (A, RV) 6.3 7.7
High 5.0 V (A, RV) 4.4 6.0
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-01-29
CE Approval Date
Registered USA Implants1109
Estimated Active USA Implants95
Normal Battery Depletions158
US Market Release1999-01-29
CE Approval Date
Estimated WW Distribution1693
Normal Battery Depletions224
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 2
Electrical Interconnect 2
Information

Malfunctions (WW)

Therapy Function Not Compromised 0
Therapy Function Compromised 3
Electrical Component 1
Electrical Interconnect 2
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050