KD701 Kappa 700 DR
Implantable Pulse Generator (IPG)
KD701 Kappa 700 DR
Information

Device Survival Probability

KD701_KD703_KD706_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr at 95.0 mo    
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0    
% Including Normal Battery Depletion 100.0 100.0 100.0 99.0 98.0 96.2 94.9 89.6    
#   289 260 229 206 186 162 137 100    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDD
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PHK
Mass 27.1 g
Volume 12.1 cc
Waveform N/A
X-Ray ID PHK
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.7 8.6
Nominal 3.5 V (A, RV) 6.3 7.7
High 5.0 V (A, RV) 4.4 6.0
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release1999-01-29
CE Approval Date1998-03-20
Registered USA Implants242
Estimated Active USA Implants47
Normal Battery Depletions21
US Market Release1999-01-29
CE Approval Date1998-03-20
Estimated WW Distribution20913
Normal Battery Depletions129
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 1
Electrical Component 1
Therapy Function Compromised 45
Electrical Component 1
Electrical Interconnect 44
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI respectively (VOO/65 with magnet).

Data as of February 9, 2015

www.medtronic.com/productperformance 351052