EMDR01 EnRhythm MRI
Implantable Pulse Generator (IPG)
EMDR01 EnRhythm MRI
Information

Device Survival Probability


EMDR01_EMDR02_SURV
    1 yr 2 yr 3 yr at 41 mo            
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0            
% Including Normal Battery Depletion 100.0 100.0 100.0 100.0            
#   155 147 120 100            
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDRP
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PTA
Mass 21.5 g
Volume 12.7 cc
Waveform N/A
X-Ray ID PTA and MRI symbol
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 9.7 11.2
Nominal 3.5 V (A, RV) 7.3 9.4
High 5.0 V (A, RV) 4.9 7.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOS Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release
CE Approval Date2008-09-30
Registered USA Implants110
Estimated Active USA Implants83
Normal Battery Depletions
US Market Release
CE Approval Date2008-09-30
Estimated WW Distribution5147
Normal Battery Depletions
Information

Malfunctions (USA)

Therapy Function Not Compromised 6
Battery 6
Therapy Function Compromised 0
Information

Malfunctions (WW)

Therapy Function Not Compromised 62
Battery 49
Electrical Component 2
Possible Early Battery Depletion 1
Other 10
Therapy Function Compromised 3
Electrical Component 1
Possible Early Battery Depletion 2
Information

Recommended Replacement (RRT)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013