E2DR31 EnPulse 2 DR
Implantable Pulse Generator (IPG)
E2DR31 EnPulse 2 DR
Information

Device Survival Probability


E2DR31_E2DR33_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 99 mo  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0  
% Including Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 99.6 98.8 98.4 98.4  
#   1398 1364 1325 1283 1180 1118 735 213 109  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PNL
Mass 31.3 g
Volume 13.1 cc
Waveform N/A
X-Ray ID PNL
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 9.0 10.1
Nominal 3.5 V (A, RV) 7.4 9.1
High 5.0 V (A, RV) 5.2 7.1
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2004-02-20
CE Approval Date2003-09-08
Registered USA Implants585
Estimated Active USA Implants393
Normal Battery Depletions8
US Market Release2004-02-20
CE Approval Date2003-09-08
Estimated WW Distribution17121
Normal Battery Depletions69
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 2
Electrical Component 1
Other 1
Therapy Function Compromised 0
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013