E2DR21 EnPulse DR
Implantable Pulse Generator (IPG)
E2DR21 EnPulse DR
Information

Device Survival Probability

E2DR21_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr at 87.0 mo    
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0    
% Including Normal Battery Depletion 99.1 98.6 97.6 94.8 87.2 59.5 15.7 3.7    
#   10178 9055 8062 6959 5660 3267 637 182    
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PMU
Mass 23.6 g
Volume 11.1 cc
Waveform N/A
X-Ray ID PMU
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 5.4 6.0
Nominal 3.5 V (A, RV) 4.3 5.4
High 5.0 V (A, RV) 3.0 4.2
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2004-02-20
CE Approval Date2003-09-08
Registered USA Implants12201
Estimated Active USA Implants1058
Normal Battery Depletions2318
US Market Release2004-02-20
CE Approval Date2003-09-08
Estimated WW Distribution16882
Normal Battery Depletions2341
Information

Malfunctions (USA)

Therapy Function Not Compromised 0
Therapy Function Compromised 1
Electrical Component 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 0
Therapy Function Compromised 1
Electrical Component 1
Information

Elective Replacement (ERI)

Telemetry indication. Rate and Mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of February 3, 2017

www.medtronic.com/productperformance 476050