E2DR01 EnPulse 2 DR
Implantable Pulse Generator (IPG)
E2DR01 EnPulse 2 DR
Information

Device Survival Probability


E2D01_E2D03_E2DR01_E2DR03_E2DR06_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 101 mo  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0  
% Including Normal Battery Depletion 99.9 99.9 99.8 99.4 98.7 97.2 91.5 70.5 13.0  
#   94812 87623 80535 73782 67131 60293 38841 9558 246  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PNB
Mass 27.1 g
Volume 12.2 cc
Waveform N/A
X-Ray ID PNB
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.5 8.5
Nominal 3.5 V (A, RV) 6.2 7.6
High 5.0 V (A, RV) 4.4 5.9
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2004-02-20
CE Approval Date2003-09-08
Registered USA Implants97119
Estimated Active USA Implants43629
Normal Battery Depletions5486
US Market Release2004-02-20
CE Approval Date2003-09-08
Estimated WW Distribution141242
Normal Battery Depletions5555
Information

Malfunctions (USA)

Therapy Function Not Compromised 21
Electrical Component 17
Possible Early Battery Depletion 1
Other 3
Therapy Function Compromised 5
Battery 1
Electrical Component 3
Electrical Interconnect 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 31
Electrical Component 23
Possible Early Battery Depletion 2
Other 6
Therapy Function Compromised 8
Battery 1
Electrical Component 6
Electrical Interconnect 1
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013

www.medtronic.com/productperformance 251050