E2DR01 EnPulse 2 DR
Implantable Pulse Generator (IPG)
E2DR01 EnPulse 2 DR
Information

Device Survival Probability

E2D01_E2D03_E2DR01_E2DR03_E2DR06_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 106.0 mo  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0  
% Including Normal Battery Depletion 99.9 99.8 99.7 99.3 98.6 97.0 91.1 71.7 16.4  
#   94881 87687 80618 73900 67280 60796 52249 24499 817  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PNB
Mass 27.1 g
Volume 12.2 cc
Waveform N/A
X-Ray ID PNB
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.5 8.5
Nominal 3.5 V (A, RV) 6.2 7.6
High 5.0 V (A, RV) 4.4 5.9
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2004-02-20
CE Approval Date2003-09-08
Registered USA Implants97065
Estimated Active USA Implants33200
Normal Battery Depletions10791
US Market Release2004-02-20
CE Approval Date2003-09-08
Estimated WW Distribution141245
Normal Battery Depletions10883
Information

Malfunctions (USA)

Therapy Function Not Compromised 20
Electrical Component 18
Possible Early Battery Depletion 1
Other 1
Therapy Function Compromised 5
Battery 1
Electrical Component 3
Electrical Interconnect 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 29
Electrical Component 24
Possible Early Battery Depletion 2
Other 3
Therapy Function Compromised 8
Battery 1
Electrical Component 6
Electrical Interconnect 1
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of January 31, 2014

www.medtronic.com/productperformance 301050