E1DR21 EnPulse DR
Implantable Pulse Generator (IPG)
E1DR21 EnPulse DR
Information

Device Survival Probability


E1DR21_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr at 80 mo      
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0      
% Including Normal Battery Depletion 100.0 99.6 98.6 96.2 91.9 61.3 23.4      
#   1640 1486 1325 1160 966 463 107      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style IS-1 UNI/BI
Serial Number Prefix PPT
Mass 23.6 g
Volume 11.1 cc
Waveform N/A
X-Ray ID PPT
Hide this content
IS-1 UNI/BI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 5.4 6.0
Nominal 3.5 V (A, RV) 4.3 5.4
High 5.0 V (A, RV) 3.0 4.2
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

From the point that the ERI is set, the pacemaker will operate at ERI conditions for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2003-12-18
CE Approval Date
Registered USA Implants1852
Estimated Active USA Implants155
Normal Battery Depletions373
US Market Release2003-12-18
CE Approval Date
Estimated WW Distribution1888
Normal Battery Depletions373
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Elective Replacement (ERI)

Telemetry indication. Rate and mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of March 12, 2013