E1DR06 EnPulse DR
Implantable Pulse Generator (IPG)
E1DR06 EnPulse DR
Information

Device Survival Probability

E1DR01_E1DR03_E1DR06_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 117.0 mo                      
% Excluding Normal Battery Depletion 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0 100.0                      
% Including Normal Battery Depletion 99.6 99.5 99.1 98.2 97.2 95.8 86.5 62.3 28.5 5.3                      
#   6002 5549 5101 4635 4200 3759 3069 1976 777 138                      
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Advisories For This Model

  • There are no advisories for this model.

Performance Notes For This Model

Information

Product Characteristics

Generator Type IPG
NBD Code DDDR
Max. Delivered Energy N/A
More
Connector Style 5/6mm UNI
Serial Number Prefix PRE
Mass 28.5
Volume 14.2
Waveform N/A
X-Ray ID PRE
Hide this content
5/6mm UNI
Information

Estimated Longevity

Amplitude Setting 500 Lead Ω 1000 Lead Ω
Low 2.5 V (A, RV) 7.5 8.5
Nominal 3.5 V (A, RV) 6.2 7.6
High 5.0 V (A, RV) 4.4 5.9
Footnote

Longevity estimates based on the following device usage. Pace/Sense Mode DDD ; Atrial Pulse Width 0.4 ms; Right Ventricle Pulse Width 0.4 ms; Lower Pace Rate 60 bpm; Right Ventricle Percent Paced 100 %; Upper Sensing Rate 130 ppm; Atrial Percent Paced 100 %; Atrial Fib Percent of Time 0 %Hide this content

Information

EOL Indication

3 months after ERI.

Information

Distribution Data

US Market Release2003-12-18
CE Approval Date
Registered USA Implants
Estimated Active USA Implants
Normal Battery Depletions
US Market Release2003-12-18
CE Approval Date
Estimated WW Distribution
Normal Battery Depletions
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Elective Replacement (ERI)

Telemetry indiction. Rate and mode change to 65 ppm and VVI (VOO/65 with magnet).

Data as of February 3, 2017

www.medtronic.com/productperformance 476050