DVMB2D4 Evera MRI XT
Implantable Cardioverter Defibrillator (ICD)
DVMB2D4 Evera MRI XT
Information

Device Survival Probability

EVERA_VR_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 106.0 mo  
% Excluding Normal Battery Depletion 100.0 100.0 99.9 99.9 99.8 99.7 99.6 99.5 99.4  
% Including Normal Battery Depletion 100.0 99.9 99.7 99.6 99.4 99.1 99.0 98.7 98.4  
#   52255 48555 45127 42009 38712 32984 19190 7315 331  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code VVE-VVIR
Max. Delivered Energy 36 J
More
Connector Style DF4 VR Style
Serial Number Prefix PLZ
Mass 77 g
Volume 33 cc
Waveform Biphasic
X-Ray ID PFZ
Hide this content
DF4 VR Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.9 5.3 5.7 5.8
Quarterly 7.8 8.8 9.7 10.1
Semi-annual 8.5 9.7 10.7 11.3
Footnote

Longevity estimates based on the following device usage. Pacing Mode VVI; Ventricular Lower Pacing Rate 60ppm; Ventricular Pulse Width 0.4ms; Ventricular Pulse Amplitude 3V; Ventricle Impedance 510ohms; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formationHide this content

Information

Distribution Data

US Market Release
CE Approval Date2014-03-31
Registered USA Implants2
Estimated Active USA Implants0
Normal Battery Depletions
US Market Release
CE Approval Date2014-03-31
Estimated WW Distribution26435
Normal Battery Depletions2
Information

Malfunctions (USA)

Therapy Function Not Compromised
Therapy Function Compromised
Information

Malfunctions (WW)

Therapy Function Not Compromised 38
Battery 28
Electrical Component 10
Therapy Function Compromised 24
Battery 22
Other 2
Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than 2.73V on 3 consecutive daily automatic measurments.

Information

Charge Time

BLACKWELL_VR_CHRG
Loading...