Device Survival Probability
BRAVA_VIVA_SURV
1 yr | 2 yr | 3 yr | 4 yr | 5 yr | 6 yr | 7 yr | 8 yr | at 101.0 mo | |||
---|---|---|---|---|---|---|---|---|---|---|---|
% | ![]() |
100.0 | 100.0 | 99.9 | 99.9 | 99.9 | 99.8 | 99.8 | 99.8 | 99.8 | |
% | ![]() |
99.9 | 99.7 | 98.8 | 96.4 | 91.6 | 82.6 | 68.0 | 48.5 | 22.2 | |
# | 86474 | 78556 | 70556 | 61290 | 49285 | 32884 | 15812 | 3785 | 103 |
- Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
- Excluding Normal Battery Depletion – This is the malfunction free survival curve.

Product Characteristics
Generator Type | CRT-Defib |
---|---|
NBD Code | DDE-DDDR |
Max. Delivered Energy | 36 J |
Connector Style | DF4 CRT-D Style |
---|---|
Serial Number Prefix | BLN |
Mass | 80 g |
Volume | 35 cc |
Waveform | Biphasic |
X-Ray ID | PXT |

Estimated Longevity
Charging Frequency | 100% Pacing | 50% Pacing | 15% Pacing | 100% Sensing |
---|---|---|---|---|
Monthly | 3.6 | 4.3 | 4.9 | 5.3 |
Quarterly | 4.9 | 6.3 | 7.8 | 8.8 |
Semi-annual | 5.2 | 6.6 | 8.5 | 9.6 |
Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Integrated bi-ventricular pacing RV+LV; Left Ventricle Impedance 510ohms; Left Ventricular Pulse Width 0.4ms; Left Ventricular Pulse Amplitude 3V; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Distribution Data
US Market Release | 2013-01-29 |
---|---|
CE Approval Date | |
Registered USA Implants | 8822 |
Estimated Active USA Implants | 2553 |
Normal Battery Depletions | 1308 |
US Market Release | 2013-01-29 |
---|---|
CE Approval Date | |
Estimated WW Distribution | 10644 |
Normal Battery Depletions | 1308 |

EOS Indication
From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.