DDMB1D1 Evera MRI XT
Implantable Cardioverter Defibrillator (ICD)
DDMB1D1 Evera MRI XT
Information

Device Survival Probability

EVERA_DR_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 99.9 99.9 99.8 99.8 99.7 99.7  
% Including Normal Battery Depletion 99.9 99.9 99.8 99.7 99.4 98.9 97.5 94.5 90.8  
#   197897 171084 138453 106095 76189 49440 28267 11880 120  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code DDE-DDDR
Max. Delivered Energy 36 J
More
Connector Style DR Style
Serial Number Prefix CWA
Mass 77 g
Volume 33 cc
Waveform Biphasic
X-Ray ID PFZ
Hide this content
DR Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.1 4.6 5.1 5.4
Quarterly 5.9 7.0 8.3 9.0
Semi-annual 6.2 7.6 9.0 9.8
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formationHide this content

Information

Distribution Data

US Market Release2016-10-12
CE Approval Date
Registered USA Implants42715
Estimated Active USA Implants34422
Normal Battery Depletions43
US Market Release2016-10-12
CE Approval Date
Estimated WW Distribution44600
Normal Battery Depletions43
Information

Malfunctions (USA)

Therapy Function Not Compromised 24
Battery 14
Electrical Component 6
Electrical Interconnect 2
Other 2
Therapy Function Compromised 7
Battery 4
Electrical Component 3
Information

Malfunctions (WW)

Therapy Function Not Compromised 27
Battery 14
Electrical Component 9
Electrical Interconnect 2
Other 2
Therapy Function Compromised 7
Battery 4
Electrical Component 3
Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than 2.73V on 3 consecutive daily automatic measurements.

Information

Charge Time

BLACKWELL_DR_CHRG
Loading...

Data as of July 27, 2022

www.medtronic.com/productperformance 876050