DDBB1D1 Evera XT
Implantable Cardioverter Defibrillator (ICD)
DDBB1D1 Evera XT
Information

Device Survival Probability

EVERA_DR_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr at 99.0 mo  
% Excluding Normal Battery Depletion 100.0 100.0 100.0 99.9 99.9 99.8 99.8 99.8 99.8  
% Including Normal Battery Depletion 99.9 99.9 99.8 99.7 99.4 98.9 97.4 94.7 94.3  
#   191123 156908 121833 89294 59636 35643 17232 2766 323  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type ICD
NBD Code DDE-DDDR
Max. Delivered Energy 36 J
More
Connector Style DR Style
Serial Number Prefix BWC
Mass 77 g
Volume 33 cc
Waveform Biphasic
X-Ray ID PXT
Hide this content
DR Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 4.1 4.6 5.1 5.4
Quarterly 5.9 7.0 8.3 9.0
Semi-annual 6.2 7.6 9.0 9.8
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2013-04-03
CE Approval Date
Registered USA Implants82180
Estimated Active USA Implants49328
Normal Battery Depletions656
US Market Release2013-04-03
CE Approval Date
Estimated WW Distribution84419
Normal Battery Depletions656
Information

Malfunctions (USA)

Therapy Function Not Compromised 64
Battery 38
Electrical Component 22
Other 4
Therapy Function Compromised 58
Battery 48
Electrical Component 5
Electrical Interconnect 2
Other 3
Information

Malfunctions (WW)

Therapy Function Not Compromised 65
Battery 38
Electrical Component 22
Other 5
Therapy Function Compromised 58
Battery 48
Electrical Component 5
Electrical Interconnect 2
Other 3
Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.73V on 3 consecutive daily automatic measurments.

Information

Charge Time

BLACKWELL_DR_CHRG
Loading...

Data as of October 11, 2021

www.medtronic.com/productperformance 826050