D314DRG Protecta XT DR
Implantable Cardioverter Defibrillator (ICD)
D314DRG Protecta XT DR
Information

Device Survival Probability

D3xxDRx_SURV
Loading...










    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr at 136.0 mo                  
% Excluding Normal Battery Depletion 100.0 99.9 99.9 99.9 99.9 99.8 99.7 99.7 99.6 99.6 99.6 99.6                  
% Including Normal Battery Depletion 99.8 99.7 99.5 99.0 97.2 91.3 77.6 52.8 38.7 37.1 36.5 36.2                  
#   54184 50301 46252 42263 37893 31012 20451 9299 5158 4196 1292 127                  
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Customer Communications For This Model

  • There are no communications for this model.
Information

Product Characteristics

Generator Type ICD
NBD Code DDE-DDDR
Max. Delivered Energy 35 J
More
Connector Style DR Style
Serial Number Prefix PSK
Mass 68 g
Volume 37 cc
Waveform Biphasic
X-Ray ID PSI
Hide this content
DR Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 3.6 4.1 4.5 4.7
Quarterly 5.1 6.0 7.0 7.5
Semi-annual 5.7 7.0 8.3 9.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

Distribution Data

US Market Release2011-03-25
CE Approval Date
Registered USA Implants34745
Estimated Active USA Implants4672
Normal Battery Depletions4544
US Market Release2011-03-25
CE Approval Date
Estimated WW Distribution35499
Normal Battery Depletions4546
Information

Malfunctions (USA)

Therapy Function Not Compromised 40
Battery 8
Electrical Component 26
Electrical Interconnect 1
Possible Early Battery Depletion 4
Other 1
Therapy Function Compromised 37
Battery 30
Electrical Component 7
Information

Malfunctions (WW)

Therapy Function Not Compromised 41
Battery 8
Electrical Component 26
Electrical Interconnect 1
Possible Early Battery Depletion 4
Other 2
Therapy Function Compromised 37
Battery 30
Electrical Component 7
Information

EOL Indicator

The Prolonged Service Period (PSP) is the time between the RRT and EOS. The PSP is defined as 3 months assuming the following conditions: 100% DDD pacing at 60 min-1, 2.5 V atrial and RV pacing amplitude; 0.4 ms pulse width; 600 Ω pacing load; and 6 full-energy charges. The EOS may be indicated before the end of 3 months if the device exceeds these conditions.

Information

Recommended Replacement (RRT)

Battery voltage less than or equal to 2.63V on 3 consecutive daily automatic measurements.

Information

Charge Time

D314DRx_CHRG
Loading...

Data as of December 1, 2023

www.medtronic.com/productperformance 951051