Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Longevity estimates based on the following device usage.
Pacing Mode DDD;
Atrial Lower Pacing Rate 60ppm;
Ventricular Lower Pacing Rate 60ppm;
Atrial Pulse Width 0.4ms;
Ventricular Pulse Width 0.4ms;
Atrial Pulse Amplitude 3V;
Ventricular Pulse Amplitude 3V;
Atrial Impedance 510ohms;
Ventricle Impedance 510ohms;
Atrial Sensing Rate 70bpm;
Ventricle Sensing Rate 70bpm;
EGM prestorage OFF;
Integrated bi-ventricular pacing RV+LV;
Left Ventricle Impedance 510ohms;
Left Ventricular Pulse Width 0.4ms;
Left Ventricular Pulse Amplitude 3V;
Optivol ON;
Atrial Fib percentage of time 0%;
Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation.
Hide this content
Distribution Data
US Market Release
2009-08-15
CE Approval Date
Registered USA Implants
30190
Estimated Active USA Implants
2595
Normal Battery Depletions
8021
US Market Release
2009-08-15
CE Approval Date
Estimated WW Distribution
30453
Normal Battery Depletions
8022
Malfunctions (USA)
Therapy Function Not Compromised
176
Battery
1
Electrical Component
22
Possible Early Battery Depletion
152
Software/Firmware
1
Therapy Function Compromised
11
Battery
1
Electrical Component
10
Malfunctions (WW)
Therapy Function Not Compromised
180
Battery
1
Electrical Component
26
Possible Early Battery Depletion
152
Software/Firmware
1
Therapy Function Compromised
11
Battery
1
Electrical Component
10
EOS Indication
From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.
Recommended Replacement (RRT)
Battery voltage less than 2.63 V or Charge Time greater than 16 seconds.