D274TRK Concerto II CRT-D
Cardiac Resynchronization Therapy - Defibrillation (CRT-Defib)
D274TRK Concerto II CRT-D
Information

Device Survival Probability


CONCERTO_II_SURV
    1 yr 2 yr 3 yr at 41 mo            
% Excluding Normal Battery Depletion 100.0 99.8 99.6 99.5            
% Including Normal Battery Depletion 99.9 98.6 94.4 93.8            
#   27022 19020 4711 108            
  • Including Normal Battery Depletion – This curve includes devices that have reached at least 80% of expected longevity. This curve is most representative of clinical performance and how long the device will last.
  • Excluding Normal Battery Depletion – This is the malfunction free survival curve.
Information

Product Characteristics

Generator Type CRT-Defib
NBD Code DDED
Max. Delivered Energy 35 J
More
Connector Style CRT-D Style
Serial Number Prefix PZV
Mass 68 g
Volume 38 cc
Waveform Biphasic
X-Ray ID PUG
Hide this content
CRT-D Style
Information

Estimated Longevity

Charging Frequency 100% Pacing 50% Pacing 15% Pacing 100% Sensing
Monthly 3.2 3.8 4.4 4.7
Quarterly 4.4 5.5 6.8 7.5
Semi-annual 4.8 6.2 7.9 9.0
Footnote

Longevity estimates based on the following device usage. Pacing Mode DDD; Atrial Lower Pacing Rate 60ppm; Ventricular Lower Pacing Rate 60ppm; Atrial Pulse Width 0.4ms; Ventricular Pulse Width 0.4ms; Atrial Pulse Amplitude 3V; Ventricular Pulse Amplitude 3V; Atrial Impedance 510ohms; Ventricle Impedance 510ohms; Atrial Sensing Rate 70bpm; Ventricle Sensing Rate 70bpm; EGM prestorage OFF; Integrated bi-ventricular pacing RV+LV; Left Ventricle Impedance 510ohms; Left Ventricular Pulse Width 0.4ms; Left Ventricular Pulse Amplitude 3V; Optivol ON; Atrial Fib percentage of time 0%; Charging frequency assumes maximum energy charge and includes therapy shocks and capacitor formation. Hide this content

Information

EOS Indication

From the point that the RRT is set, the pacemaker will operate for approximately three months for typical pacemaker configurations during the normal operating life.

Information

Distribution Data

US Market Release2009-08-15
CE Approval Date
Registered USA Implants30146
Estimated Active USA Implants23962
Normal Battery Depletions332
US Market Release2009-08-15
CE Approval Date
Estimated WW Distribution30452
Normal Battery Depletions334
Information

Malfunctions (USA)

Therapy Function Not Compromised 72
Electrical Component 6
Possible Early Battery Depletion 57
Software/Firmware 1
Other 8
Therapy Function Compromised 1
Electrical Component 1
Information

Malfunctions (WW)

Therapy Function Not Compromised 76
Electrical Component 10
Possible Early Battery Depletion 57
Software/Firmware 1
Other 8
Therapy Function Compromised 1
Electrical Component 1
Information

Recommended Replacement (RRT)

Battery voltage less than 2.63 V or Charge Time greater than 16 seconds.

Information

Charge Time


D274TRK_D294TRK_CHRG

Data as of March 12, 2013

www.medtronic.com/productperformance 251050